Despite the superior quality of the NiTi alloy compared to traditional stainless steel instruments, used in rotational or reciprocating movements in the root canals with significant curvature, they present a great risk of fracture, which could compromise the success of endodontic treatment1-6.
The shaping and decontamination of root canals cause superficial defects in NiTi instruments observed through Scanning Electron Microscopy (SEM). Longitudinal microcracks can be observed along the axis of the file, and distortion can be observed in machined grooves when subjected to torsional loads7. The improvement in the fatigue resistance of endodontic instruments is achieved thanks to the optimization of the microstructure of its alloy; this is achieved by thermomechanical processing8.
Another critical factor in preventing instrument fractures is to prevent them from going beyond the elastic limit of the metal alloy, which is why it has been proposed to work with new movements during instrumentation, such as reciprocal movement. It is claimed that using of reciprocating motion for a nickel-titanium file increasesfatigue strength compared to continuous rotation9.
Among the reciprocating systems that are manufactured by changing the molecular structure through a new heat treatment to increase the resistance to cyclic fatigue, and this new heat treatment gives the files their blue color, as Reciproc Blue files (VDW, Munich, Germany) which has an S-shaped cross-section, two cutting edges and a non-cut tip, variable taper 0,0810, the AF Blue R3 files (Fanta Dental, Shanghai, China) have a square cross-section and a fixed taper of 0.06 and the X1 Blue File file (MK Life, Porto Alegre, RS, Brazil) which has an inactive tip and a triangular cross-section and a fixed 0,06 taper.
Fracture strength studies of various rotary systems are essential. The aim of this in vitro study was to compare the dynamic cyclic fatigue strength of 3 reciprocating NiTi files with heat treatment Reciproc Blue (25/08), AF Blue R3 (25/06), and X1 Blue File (25/06).
A total of 30 new endodontic files, AF Blue R3 Blue 25/06 (Fanta Dental, Shanghai, China), X1 Blue File 25/06 (MK Life, Porto Alegre, RS, Brazil), Reciproc Blue 25/08 (VDW, Munich, Germany), were selected for this experiment. They were from two different production lots. Each file was examined using an operating microscope (Alliance, São Carlos, São Paulo, Brazil) with a magnification of 24 to verify that it does not have defects and deformities.
The sample size calculation was carried out with the EPIDAT program with the following data: significance level 5%, power of 90%, mean differences of 20 (standard deviations of 15,3 and 11,2). The sample size calculation indicates the need for the present study to have 10 files in each group.
For dynamic cyclic fatigue resistance tests, a groove was made simulating an artificial root canal in a stainless steel plate (Figure 1), according to criteria established in previous studies11,12. The curved metal artificial root canal had an angle of 69° and a radius of 2,5 mm. The straight part of the cervical canal up to the limit preceding the beginning of the curvature was 13,35 mm long, the length of the curvature was 2,15 mm, and that of the straight part of the posterior canal at the end of the curvature to the apex it had a length of 7,5 mm. The width of the canal of the cervical part was 2 mm, at the limit before the curvature of 1.52 mm, at the final limit of the curvature was 1.49 mm, and the apical width was 1.00 mm. The purporse of these dimensions was to allow the freely rotation of the file within the canal. An acrylic plate was fixed on the stainless-steel plate so there was no loss of the separated fragment, and the file was contained within the simulated canal during the experiment.
The instruments were coupled to the 6:1 contra-angle (Sirona Dental Systems GmbH, Benshein, Germany), of a silver VDW engine (Dentisply Industria e Comércio Ltda; VDW GmbHMunich - Germany) the files were activated using the programming "RECIPROC ALL." To reproduce the axial movements of the instruments that occur during a clinical procedure, the reducing contra-angle was coupled to a mechanical system of vertical movement to avoidvariations in position and angulation. The tests were performed dynamically, where the mechanical system of vertical movement was driven by a SAVOX SC-12 56T69 engine (Savox All Rights Reserved, Taichung, Taiwan) which makes the mechanical system penetrate and traction what is done through a linear guide, with servo movement controlled by an electronic processor that controls the speed and amplitude of penetration and traction.
With the endodontic contra-angle disabled, the files were inserted 22 mm in the artificial canal and activated in a reciprocating movement to the left and underwent a penetration and axial traction movement with an amplitude of 3 mm in penetration movement and 3 mm in the direction of traction, the penetration and traction movement starts simultaneously with the triggering of the reciprocating movement of the file within the canal.
During the experiment, the artificial root canal was filled with lubricating oil (Synger, La Vergne, Tennessee, USA) using a 24G 3/4 (0,55 x 20 mm) hypodermic needle (Shandong Weigao Group Medical Polymer, Shandong Province, China). The endodontic motor was activated, thus initiating the dynamic cyclic fatigue test, immediately the linear guidance device exerted the penetration movement of 3 mm and then performed the traction movement of 3 mm, with a speed of 3 mm / s for both movements13, standardized and until the file fractures.
During axial penetration and traction movement so, that the files never changed the range of motion and always remained with their active part within the length of the curvature, an amplitude of 3 mm in each direction was used to simulate the pecking movement used in an actual clinical situation.
The stainless steel plate with the simulated channel was placed on a heated plate of a magnetic stirrer at a controlled temperature of 35oC ± 1oC better simulate the intrachannel temperature conditions14, and confirmation of this temperature was performed by a digital infrared thermometer (G-Tech) before starting the test each.
Video recording was carried out simultaneously, from the beginning of the file activation until its fracture, with a digital cell phone camera (Huawei p30, Shenzhen, Guangzhou, China) to avoid possible errors and allow checking the time until the fracture occurs. The video was transferred to a computer and using Windows Media Player software (Microsoft Windows, Redmond, Washington, USA) which allowed us to analyze the exact moment when the file is activated within the simulated canal until the moment of fracture, considered a time for fracture of each file6,15.
All fractured fragments of the samples underwent a cleaning phase in an ultrasonic vat and dried in an oven at 37oC for 24 hours, placed in plastic tubes identified with the name and number of the instrument. They were then analyzed using a scanning electron microscope (SEM; EVO 15; ZEISS, Cambridge, UK), where the separate file fragment was photographed horizontally and transversely.
Fractured file fragments were measured with a 150 mm digital gauge, precision ± 0,03 mm / 0,001 (hardened stainless steel, model MTX 316119). And the following were analyzed: the pattern in the fracture length of the files in each group and the location of the maximum tension point in the artificial test canal.
Statistical analysis was performed using the Stata 15,0 Software. The Shapiro-Wilk test was applied to analyze normality for both the time variable and mm variables. Considering a significance level of 5%, all variables were normally distributed. The ANOVA test was applied for theanalysis of multiple groups. Tukey's test was then used for intergroup analysis.
Each of the groups consisted of 10 and 25 mm files. The procedure was performed at controlled temperatures (Table 1). The onset temperature was 34,9 ±0,3 in the AFB group. The temperature during the process was 35,6 ±1,3 in the AFB group. The mm of the fragment fracture was 6,5 ±0,5 in the AFB group.
Start temperature | Temperature during the procedure | Fracture Fragment mm | |
---|---|---|---|
X1B | 35,1 ±0,3 | 35,1 ±0,3 | 7,1 ±0,5 |
AFB | 34,9 ±0,3 | 35,6 ±1,3 | 6,5 ±0,5 |
RB | 35,1 ±0,3 | 35,2 ±0,4 | 6,5 ±0,6 |
The mean time in seconds until the file fractured was 170,7 ±15,1 in the AFB group. This difference was statistically significant (p: 0,000) compared to the X1B and RB group with the ANOVA test (Table 2). Subsequently, with the tukey test, it was demonstrated that these differences are presented by comparing X1B with AFB (p: 0,000) and X1B with RB (p: 0,000). However, there are no statistically significant differences between RB and AFB (p:0,739).
Tempo | Mean | SD | min-max | p-value |
---|---|---|---|---|
X1B | 110,4b | 26,8 b | 55-148 b | 0,0000 |
AFB | 170,7a | 15,1 a | 149-197 a | |
RB | 163,3a | 22,9 a | 122-195 a |
Crack initiation areas and overload rapid fracture zones were also observed without morphological differences in the fracture surfaces of all evaluated instruments. The micro-cavities observed on the instrument surfaces evolve from the first test cycles, producing micro-cracks in AFB, X1B, and RB files, respectively (Figure 2). Often, these microcracks come together to create cracks, usually in the axial direction. However, the failure only occurs when this defect grows in or near the file.
The Scanning electron microscopy images (SEM) analysis of the instruments submitted to the cyclic fatigue resistance test showed typical characteristics of cyclic fatigue fracture, that is, morphological characteristics of the ductile type with numerous dimples in AFB, X1B, and RB files, respectively (Figure 3). The images show multiple grooves scattered across the fractured surfaces, which are a typical feature of ductile fracture.
In this studio, 30 new files from the RB, AFB, and X1B brands were tested because the files are manufactured from the same NiTi alloy and have the same surface heat treatment. The instrument sizes of 10 and 25 mm was tested because these sizes are commonly used during instrumentation.
The additional heat that NiTi instruments are subjected to during autoclave sterilization may vary their CF resistance. This was shown in a study that found that fatigue resistanceis higher in new files16. In this study, all files used were new.
The methodology used in this study allowed the evaluation of dynamic cyclic fatigue resistance at intracanal temperature 35ºC ±1ºC. This temperature was observed in other studies6,14 because it reproduces more accuratly what happens inside the canal during instrumentation.
The curved metallic artificial channel used in this studio had an angle of 69º and a 2,5mm radius to ensure standardization of the experimental conditions and minimize any factors that might affect the results17-19.
In addition, in the present study, a dynamic model was used to simulate better systems`s clinical use of the, in which the axial movement was fixed at 3 mm20.
This study found differences in time in seconds until file fractures and NCF between X1B compared to AFB and RB. However, there are no differences between AFB and RB.
The cross-section is not the only parameter influencing the fatigue resistance between AFB, X1B, and RB files. Also, the diameter of RB is diferente than others. Therefore, an explanation for the longer service life of the RB and AFB of the X1B could be related to the different cross-sections of the two instruments. RB has an S-shaped cross-section, while X1B has a triangular cross-section. A triangular cross-section design with three cutting edgescould increase the file`s torsional stress on the by decreasing its cyclic fatigue resistance21. The S-shaped cross section can also be advantageous in reducing torsional stresses due to the smaller surface area in contact with the channel walls. In addition, the S-shaped cross-section design exhibits less blade interaction thanto the convex triangular cross-section and can influence the cyclic fatigue resistance of the instrument3. The thermomechanical CM treatment gives greater resistance to cyclic fatigue23. In addition, CM wire instruments have greater flexibility than conventional M-Wire and NiTi instruments23-27.
Crack initiation areas and overload rapid fracture zones without morphological differences in the fracture surfaces of all evaluated instruments were also observed. The SEM analysis of the instruments submitted to the cyclic fatigue resistance test showed typical characteristics of cyclic fatigue fracture, that is, morphological characteristics of the ductile type with numerous dimples.
Regarding length of the broken segment of the file not diferente on our study, diferente of another study that showed significant difference in length of fractured fragments of four instruments (Reciproc R25, One Curve, Fanta-one, and Fanta R3- Reciproc). The one curve systems showed the shortest length of fractured fragments, probably because of the differences of the cross section along the file length. Fracture of the instruments occurred at the center or below the curvature which affirms the importance of the positioning of the files in the artificial root canal28.
Regarding the limitations of the in vitro study, it was conducted under strictly controlled conditions without giving advantage to any study group. The curved simulated flute dynamic model standardizes variables affecting cyclic fatigue resistance, such as flute diameter, canal wall, and canal inlet length so that the file would rotate freely within the canal. Fanta Dental and the the X1 Blue files are made with the same NiTi and heat treatment. difference between these two files is the cross sections. It is well known that the cyclic fatigue resistance is more significant when the core is smaller. All things equal, a triangular cross-section will result in less time to fracture compared to a quadrangular cross-section. Even thoungh we calculated the number of cycles to fracture; it was not valid for reciprocating movements. Therefore, those data were not valid and were removed. As a limitation, it can be mentioned that different brands are not compared, so the need to carry out other studies arises. As a strength, a sample size calculation was performed to boost the study.
Very few studies were found comparing the strength of AFB against other files. Further experimental research or randomized clinical trials with a significant number of samples are recommended to find more conclusive results, considering the cost of AFB, which is much more affordable than RB.
1. Calderón Gutiérrez AK. Evaluación de la resistencia a la fatiga cíclica de diferentes instrumentos rotatorios: Protaper Gold, Protaper Next, Hyflex CM, Reciproc, Wave One y Wave One Gold. [Thesis]. México: Universidad Autónoma de Nuevo León; 2016. Available in http://eprints.uanl.mx/14197/1/1080237922.pdf
AK. Calderón Gutiérrez Evaluación de la resistencia a la fatiga cíclica de diferentes instrumentos rotatorios: Protaper Gold, Protaper Next, Hyflex CM, Reciproc, Wave One y Wave One Gold.ThesisMéxicoUniversidad Autónoma de Nuevo León2016http://eprints.uanl.mx/14197/1/1080237922.pdf
2. De-Deus G, Leal Vieira VT, Nogueira da Silva EJ, Lopes H, Elias CN, Moreira EJ. Bending resistance and dynamic and static cyclic fatigue life of Reciproc and WaveOne large instruments. J Endod. 2014; 40(4): 575-9. DOI: https://doi.org/10.1016/j.joen.2013.10.013
G De-Deus VT Leal Vieira EJ Nogueira da Silva H Lopes CN Elias EJ Moreira Bending resistance and dynamic and static cyclic fatigue life of Reciproc and WaveOne large instrumentsJ Endod.2014404575579https://doi.org/10.1016/j.joen.2013.10.013
3. Di Nardo D, Gambarini G, Seracchiani M, Mazzoni A, Zanza A, Giudice A, et al. Influence of different cross-section on cyclic fatigue resistance of two nickel-titanium rotary instruments with same heat treatment: an in vitro study. Saudi Endod J. 2020; 10(3): 221-5. DOI: http://dx.doi.org/10.4103/sej.sej_124_19
D Di Nardo G Gambarini M Seracchiani A Mazzoni A Zanza A Giudice Influence of different cross-section on cyclic fatigue resistance of two nickel-titanium rotary instruments with same heat treatment: an in vitro studySaudi Endod J.2020103221225http://dx.doi.org/10.4103/sej.sej_124_19
4. Jamleh A, Alghaihab A, Alfadley A, Alfawaz H, Alqedairi A, Alfouzan K. Cyclic fatigue and torsional failure of edgetaper platinum endodontic files at simulated body temperature. J Endod. 2019; 45(5): 611-4. DOI: https://doi.org/10.1016/j.joen.2019.02.008
A Jamleh A Alghaihab A Alfadley H Alfawaz A Alqedairi K Alfouzan Cyclic fatigue and torsional failure of edgetaper platinum endodontic files at simulated body temperatureJ Endod.2019455611614https://doi.org/10.1016/j.joen.2019.02.008
5. Özyürek T. Cyclic fatigue resistance of reciproc, waveone, and waveone gold nickel-titanium instruments. J Endod. 2016; 42(10): 1536-9. DOI: https://doi.org/10.1016/j.joen.2016.06.019
T. Özyürek Cyclic fatigue resistance of reciproc, waveone, and waveone gold nickel-titanium instrumentsJ Endod.2016421015361539https://doi.org/10.1016/j.joen.2016.06.019
6. Topçuoğlu HS, Demirbuga S, Düzgün S, Topçuoğlu G. Cyclic fatigue resistance of new reciprocating files (Reciproc Blue, WaveOne Gold, and SmartTrack) in two different curved canals. J Investig Clin Dent. 2018; 9(3): e12344. DOI: https://doi.org/10.1111/jicd.12344
HS Topçuoğlu S Demirbuga S Düzgün G Topçuoğlu Cyclic fatigue resistance of new reciprocating files (Reciproc Blue, WaveOne Gold, and SmartTrack) in two different curved canalsJ Investig Clin Dent.201893e12344https://doi.org/10.1111/jicd.12344
7. Arboleda Gaviria AM, Huertas Valens ER. Cambios superficiales de los sistemas WaveOne Gold® y Reciproc Blue®: revisión de literatura [Thesis]. Bogotá: Universidad Santo Tomás; 2020. Available in https://repository.usta.edu.co/bitstream/handle/11634/27988/2020ArboledaAdriana.pdf?sequence=6&isAllowed=y
AM Arboleda Gaviria ER Huertas Valens Cambios superficiales de los sistemas WaveOne Gold® y Reciproc Blue®: revisión de literaturaThesisBogotáUniversidad Santo Tomás2020https://repository.usta.edu.co/bitstream/handle/11634/27988/2020ArboledaAdriana.pdf?sequence=6&isAllowed=y
8. Cortés Naranjo DG. Evaluación de la extrusión apical de barro dentinario asociado a sistemas de instrumentación rotatoria y reciprocante en conductos curvos: estudio in vitro [Thesis]. Quito: Universidad Central del Ecuador; 2019. Available in: http://www.dspace.uce.edu.ec/bitstream/25000/18569/1/T-UCE-0015-ODO-029-P.pdf
DG Cortés Naranjo Evaluación de la extrusión apical de barro dentinario asociado a sistemas de instrumentación rotatoria y reciprocante en conductos curvos: estudio in vitroThesisQuitoUniversidad Central del Ecuador2019http://www.dspace.uce.edu.ec/bitstream/25000/18569/1/T-UCE-0015-ODO-029-P.pdf
9. Lee W, Hwang YJ, You SY, Kim HC. Effect of reciprocation usage of nickel-titanium rotary files on the cyclic fatigue resistance. Aust Endod J. 2013; 39(3): 146-50. DOI: https://doi.org/10.1111/j.1747-4477.2012.00352.x
W Lee YJ Hwang SY You HC Kim Effect of reciprocation usage of nickel-titanium rotary files on the cyclic fatigue resistanceAust Endod J.2013393146150https://doi.org/10.1111/j.1747-4477.2012.00352.x
10. Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments. J Endod. 2017; 43(7): 1192-6. DOI: https://doi.org/10.1016/j.joen.2017.03.009
M Gündoğar T Özyürek Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium InstrumentsJ Endod.201743711921196https://doi.org/10.1016/j.joen.2017.03.009
11. De-Deus G, Silva EJNL, Vieira VTL, Belladonna FG, Elias CN, Plotino G, et al. Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the reciproc files. J Endod. 2017; 43(3): 462-6. DOI: http://dx.doi.org/10.1016/j.joen.2016.10.039
G De-Deus EJNL Silva VTL Vieira FG Belladonna CN Elias G Plotino J Endod.2017433462466http://dx.doi.org/10.1016/j.joen.2016.10.039
12. Kim HC, Kwak SW, Cheung GSP, Ko DH, Chung SM, Lee W. Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: reciproc versus WaveOne. J Endod. 2012; 38(4): 541-4. DOI: https://doi.org/10.1016/j.joen.2011.11.014
HC Kim SW Kwak GSP Cheung DH Ko SM Chung W Lee Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: reciproc versus WaveOneJ Endod.2012384541544https://doi.org/10.1016/j.joen.2011.11.014
13. Keskin C, Inan U, Demiral M, Keleş A. Cyclic fatigue resistance of Reciproc Blue, Reciproc, and WaveOne Gold reciprocating instruments. J Endod. 2017; 43(8): 1360-3. DOI: https://doi.org/10.1016/j.joen.2017.03.036
C Keskin U Inan M Demiral A Keleş Cyclic fatigue resistance of Reciproc Blue, Reciproc, and WaveOne Gold reciprocating instrumentsJ Endod.201743813601363https://doi.org/10.1016/j.joen.2017.03.036
14. De Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015; 41(7): 1112-5. DOI: https://doi.org/10.1016/j.joen.2015.02.011
F De Hemptinne G Slaus M Vandendael W Jacquet RJ De Moor P Bottenberg In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochloriteJ Endod.201541711121115https://doi.org/10.1016/j.joen.2015.02.011
15. Silva EJNL, Villarino LS, Vieira VTL, Accorsi-Mendonça T, Antunes HDS, De-Deus G, et al. Bending resistance and cyclic fatigue life of reciproc, unicone, and WaveOne reciprocating instruments. J Endod. 2016; 42(12): 1789-93. DOI: https://doi.org/10.1016/j.joen.2016.08.026
EJNL Silva LS Villarino VTL Vieira T Accorsi-Mendonça HDS Antunes G De-Deus Bending resistance and cyclic fatigue life of reciproc, unicone, and WaveOne reciprocating instrumentsJ Endod.2016421217891793https://doi.org/10.1016/j.joen.2016.08.026
16. Bermejo Sánchez AC. Efecto de los procesos termomecánicos en la resistencia a fatiga [Thesis]. Madrid: Universidad Complutense de Madrid; 2018. Available in https://eprints.ucm.es/id/eprint/51175/1/T40855.pdf
AC. Bermejo Sánchez Efecto de los procesos termomecánicos en la resistencia a fatigaThesisMadridUniversidad Complutense de Madrid2018https://eprints.ucm.es/id/eprint/51175/1/T40855.pdf
17. Al-Obaida MI, Merdad K, Alanazi MS, Altwaijry H, AlFaraj M, Alkhamis AA, et al. Comparison of cyclic fatigue resistance of 5 heat-treated nickel-titanium reciprocating systems in canals with single and double curvatures. J Endod. 2019; 45(10): 1237-41. DOI: https://doi.org/10.1016/j.joen.2019.06.011
MI Al-Obaida K Merdad MS Alanazi H Altwaijry M AlFaraj AA Alkhamis Comparison of cyclic fatigue resistance of 5 heat-treated nickel-titanium reciprocating systems in canals with single and double curvaturesJ Endod.2019451012371241https://doi.org/10.1016/j.joen.2019.06.011
18. Thu M, Ebihara A, Maki K, Miki N, Okiji T. Cyclic Fatigue resistance of rotary and reciprocating Nickel-Titanium instruments subjected to static and dynamic tests. J Endod. 2020; 46(11): 1752-7. DOI: https://doi.org/10.1016/j.joen.2020.08.006
M Thu A Ebihara K Maki N Miki T Okiji Cyclic Fatigue resistance of rotary and reciprocating Nickel-Titanium instruments subjected to static and dynamic testsJ Endod.2020461117521757https://doi.org/10.1016/j.joen.2020.08.006
19. Pedullà E, La Rosa GRM, Virgillito C, Rapisarda E, Kim HC, Generali L. Cyclic fatigue resistance of Nickel-titanium rotary instruments according to the angle of file access and radius of root canal. J Endod. 2020; 46(3): 431-6. DOI: https://doi.org/10.1016/j.joen.2019.11.015
E Pedullà GRM La Rosa C Virgillito E Rapisarda HC Kim L Generali Cyclic fatigue resistance of Nickel-titanium rotary instruments according to the angle of file access and radius of root canalJ Endod.2020463431436https://doi.org/10.1016/j.joen.2019.11.015
20. Keleş A, Eymirli A, Uyanık O, Nagas E. Influence of static and dynamic cyclic fatigue tests on the lifespan of four reciprocating systems at different temperatures. Int Endod J. 2019; 52(6): 880-6. DOI: https://doi.org/10.1111/iej.13073
A Keleş A Eymirli O Uyanık E Nagas Influence of static and dynamic cyclic fatigue tests on the lifespan of four reciprocating systems at different temperaturesInt Endod J.2019526880886https://doi.org/10.1111/iej.13073
21. Ismail AG, Galal M, Nagy MM. Effect of different kinematics and operational temperature on cyclic fatigue resistance of rotary NiTi systems. Bull Natl Res Cent. 2020; 44(1): 116.
AG Ismail M Galal MM Nagy Effect of different kinematics and operational temperature on cyclic fatigue resistance of rotary NiTi systemsBull Natl Res Cent.2020441116116
22. Zupanc J, Vahdat-Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys - a review. Int Endod J. 2018; 51(10): 1088-103. DOI: https://doi.org/10.1111/iej.12924
J Zupanc N Vahdat-Pajouh E Schäfer New thermomechanically treated NiTi alloys - a reviewInt Endod J.2018511010881103https://doi.org/10.1111/iej.12924
23. Ninan E, Berzins DW. Torsion and bending properties of shape memory and superelastic Nickel-Titanium rotary instruments. J Endod. 2013; 39(1): 101-4. DOI: https://doi.org/10.1016/j.joen.2012.08.010
E Ninan DW Berzins Torsion and bending properties of shape memory and superelastic Nickel-Titanium rotary instrumentsJ Endod.2013391101104https://doi.org/10.1016/j.joen.2012.08.010
24. Pereira GKR, Guilardi LF, Dapieve KS, Kleverlaan CJ, Rippe MP, Valandro LF. Mechanical reliability, fatigue strength and survival analysis of new polycrystalline translucent zirconia ceramics for monolithic restorations. J Mech Behav Biomed Mater. 2018; 85: 57-65. DOI: https://doi.org/10.1016/j.jmbbm.2018.05.029
GKR Pereira LF Guilardi KS Dapieve CJ Kleverlaan MP Rippe LF Valandro Mechanical reliability, fatigue strength and survival analysis of new polycrystalline translucent zirconia ceramics for monolithic restorationsJ Mech Behav Biomed Mater.2018855765https://doi.org/10.1016/j.jmbbm.2018.05.029
25. Goo HJ, Kwak SW, Ha JH, Pedullà E, Kim HC. Mechanical properties of various heat-treated Nickel-titanium rotary instruments. J Endod. 2017; 43(11): 1872-7. DOI: https://doi.org/10.1016/j.joen.2017.05.025
HJ Goo SW Kwak JH Ha E Pedullà HC Kim Mechanical properties of various heat-treated Nickel-titanium rotary instrumentsJ Endod.2017431118721877https://doi.org/10.1016/j.joen.2017.05.025
26. Soares RG, Lopes HP, Elias CN, Viera MVB, Vieira VTL, De Paula CB, et al. Comparative study of the mechanical properties of instruments made of conventional, M-wire, R-phase, and controlled memory nickel-titanium alloys. Endo: endodontic practice today. 2017; 11(4): 271-7.
RG Soares HP Lopes CN Elias MVB Viera VTL Vieira CB De Paula Comparative study of the mechanical properties of instruments made of conventional, M-wire, R-phase, and controlled memory nickel-titanium alloysEndo: endodontic practice today2017114271277
27. Santos L de A, Bahia MG de A, Las Casas EB de, Buono VTL. Comparison of the mechanical behavior between controlled memory and superelastic Nickel-Titanium files via finite element analysis. J Endod. 2013; 39(11): 1444-7. DOI: https://doi.org/10.1016/j.joen.2013.07.030
L de A Santos MG de A Bahia EB de Las Casas VTL Buono Comparison of the mechanical behavior between controlled memory and superelastic Nickel-Titanium files via finite element analysisJ Endod.2013391114441447https://doi.org/10.1016/j.joen.2013.07.030
28. Altufayli MD, Salem B, Katbeth I, Georges P, Moualla MR. Evaluation of cyclic fatigue resistance of four heat-treated nickel-titanium single file systems in canals with single curvature: an in-vitro study. Medical Science. 2021; 25(115): 2275-80.
MD Altufayli B Salem I Katbeth P Georges MR Moualla Evaluation of cyclic fatigue resistance of four heat-treated nickel-titanium single file systems in canals with single curvature: an in-vitro studyMedical Science20212511522752280
[5]Cómo citar: Portillo-Martínez MA, Horta-Dos Santos LM, Rangel-Do Couto V, Frozoni M, Méndez J. Cyclic fatigue resistance of three reciprocating nickel-titanium instruments with heat treatment at intrachannel temperature. Rev Fac Odontol Univ Antioq. 2023; 35(1). 25-35. DOI: http://dx.doi.org/10.17533/udea.rfo.v35n1a2