Voltammetric Determination of Ascorbic Acid in Pharmaceutical Formulations Using Modified Iodine-Coated Platinum Electrode


Abstract

Background:

Despite the high reactivity of the platinum electrode, the iodine-coated platinum electrode shows obvious inertness toward adsorption and surface processes. For that, iodine-coated platinum electrodes accommodate themselves to interesting voltammetric applications.

Objectives:

This study reports using the modified iodine-coated polycrystalline platinum electrode as a voltammetric sensor for ascorbic acid determination in pharmaceutical formulations.

Methods:

The developed voltammetric method based on recording cyclic voltammograms of ascorbic acid at iodine-coated electrode The optimized experimental parameters for the determination of ascorbic acid were using 0.1 M KCl as a supporting electrolyte with a scan rate of 50mV/s.

Results:

The anodic peak related to ascorbic acid oxidation was centered at nearly 0.28V. An excellent and extended linear dependence of the oxidative peak current on the concentration of ascorbic acid was observed in the range 2.84x10-3 - 5.68 mM. The limit of detection (LOD) and limit of quantitation (LOQ) were 1.0 µM and 3.01 µM, respectively, attesting to the method’s sensitivity. The investigation for the effect of potential interference from multivitamin tablet ingredients (vitamins B1, B6, B12, folic acid, citric acid, sucrose, glucose, and zinc) indicated specific selectivity toward ascorbic acid and the absence of any electrochemical response toward these components. Recovery results in the range 98.93±2.78 - 99.98±5.20 for spiked standard ascorbic acid in pharmaceutical formulations further confirmed the potential applicability of the developed method for the determination of ascorbic acid in real samples.

Conclusions:

The developed method was successfully applied to the analysis of ascorbic acid (vitamin C), and the obtained results were in good agreement with the labeled values; besides, the statistical tests indicated no significant difference at p=0.05 with a 95% confidence level.

Keywords:

ascorbic acid analysis, pharmaceutical formulation, voltammetric analysis, iodine-coated platinum electrode, modified platinum electrode


Resumen

Antecedentes:

A pesar de su alta reactividad, el electrodo de platino recubierto de yodo muestra una inercia evidente hacia la adsorción y los procesos superficiales. Por ello, los electrodos de platino recubiertos de yodo se adaptan a interesantes aplicaciones voltamétricas.

Objetivos:

Este estudio informa sobre el uso del electrodo de platino policristalino recubierto de yodo modificado como sensor voltamétrico para la determinación del ácido ascórbico en formulaciones farmacéuticas.

Métodos:

El método voltamétrico desarrollado se basa en el registro de voltamperogramas cíclicos del ácido ascórbico en el electrodo recubierto de yodo Los parámetros experimentales optimizados para la determinación del ácido ascórbico fueron utilizando KCl 0,1 M como electrolito de soporte con una velocidad de barrido de 50mV/s.

Resultados:

El pico anódico relacionado con la oxidación del ácido ascórbico se centró en casi 0,28V. Se observó una excelente y extendida dependencia lineal de la corriente del pico oxidativo con respecto a la concentración de ácido ascórbico en el rango 2,84x10-3 - 5,68 mM. El límite de detección (LOD) y el límite de cuantificación (LOQ) fueron 1,0 µM y 3,01 µM, respectivamente, lo que demuestra la sensibilidad del método. La investigación del efecto de la interferencia potencial de los ingredientes de las tabletas multivitamínicas (vitaminas B1, B6, B12, ácido fólico, ácido cítrico, sacarosa, glucosa y zinc) indicó una selectividad específica hacia el ácido ascórbico y la ausencia de cualquier respuesta electroquímica hacia estos componentes. Los resultados de recuperación en el rango de 98,93±2,78 - 99,98±5,20 para el ácido ascórbico estándar adicionado en formulaciones farmacéuticas confirmaron además la potencial aplicabilidad del método desarrollado para la determinación del ácido ascórbico en muestras reales.

Conclusiones:

El método desarrollado se aplicó con éxito al análisis de ácido ascórbico (vitamina C), y los resultados obtenidos coincidieron con los valores etiquetados; además, las pruebas estadísticas no indicaron diferencias significativas a p=0,05 con un nivel de confianza del 95%.

Palabras clave:

análisis de ácido ascórbico, formulación farmacéutica, análisis voltamétrico, electrodo de platino recubierto de yodo, electrodo de platino modificado


Introduction

L-ascorbic acid (C6H8O6) is the trivial name of Vitamin C, scheme 1, considered one of the essential water-soluble vitamins for human health. It is found in various biological systems and fresh foodstuff (1). The human body required ascorbic acid for normal physiological functions such as the synthesis and metabolism of tyrosine, folic acid, and tryptophan (2). At the same time, ascorbic acid deficiency is associated with many diseases such as anemia infections and scurvy (3).

Scheme 1

The chemical structure of ascorbic acid

0121-4004-vitae-28-02-e7-gch1.jpg

Additionally, synthetic ascorbic acid is available in several types of supplements such as tablets, capsules, chewable tablets, crystalline powder, effervescent tablets, and liquid forms (4). Several methods have been reported in the literature for the quantitative determination of ascorbic acid in various matrices. These methods include chromatography (5,6), titration (7), spectroscopy (8-12), fluorimetry (13), and flow injection analysis (14). However, some of the reported methods are time-consuming, and some are expensive and need skilled personnel.

Alternatively, electrochemical methods are considered a promised methods because of the short time response, low cost, sensitivity, and simplicity of instrumentation (15).

The modification of electrode surface is a quest to render electrochemical function that is not possible or difficult to achieve by using conventional electrodes. The goals of the improvement process include increasing selectivity, sensitivity, chemical and electrochemical stability, large usable potential window, and improving resistance to fouling (contaminating) (16).

Therefore, the avenues have opened toward the modification of solid electrodes (17). For instance, adsorbed iodine on platinum electrode surface enhances voltammetry’s reproducibility and simplifies background behavior (18). Also, coating of solid electrodes surface alters the kinetics and mechanisms of reactions run at the electrode surface. Iodine is one of the anions adsorbed to an electrode surface. The chemisorption process is achieved in two ways; from solution or vacuum to form stable chemisorbed monolayers; subsequently, the iodine-coated electrode is rinsed or evacuated (19). The iodine is adsorbed at potential 0.2 V vs. Ag/AgCl or SCE reference electrodes (20), which is the double layer potential. At the surface of the polycrystalline platinum electrode, the reaction of the iodine anions from the solution leads to spontaneous chemisorption of iodide anion to form stable neutral iodine atoms accompanied by an evolution of hydrogen gas. The adsorbed iodine is less reactive toward electrochemical oxidation than the free iodine anion in solution (21) and depends on the electrode potential (22). The chemisorbed iodine could be desorbed from the platinum electrode surface if the potential scanned is lower than -0.2 V, reducing of hydrogen ions and hydrogen gas generation (21). Also, the rate of iodine desorption from the electrode surface increases as the potential becomes more negative (22). In the positive direction, the chemisorbed iodine begins to desorb at a potential of 1.0 V (20). Studies have shown that carbon monoxide (CO) completely desorbed iodine from platinum electrode surface at potentials lower than 0.35 V, while at higher potentials, the desorption is incomplete (23).

Iodine-coated platinum electrode has been applied for the electrochemical determination of organic and non-organic species in many studies (24-32). Iodine-coated platinum electrode is characterized by the simplicity in preparation, application, and use of environmentally friendly chemical reagents. The simplicity of the method’s instrumentation stimulates our interest in this research. This work develops a simple method for ascorbic acid determination in pharmaceutical formulations.

Experimental

Materials and apparatus

A potentiostat (PAR Model 362, EG & G) interfaced to a computer via a GPIB interface (IEEE) data acquisition was used. Locally modified Labview® (IEEE) software was used for data acquisition. A one-compartment electrochemical cell with one inlet/outlet for gas purging and blanketing with oxygen-free nitrogen was used. The working electrode was a 0.5 mm polycrystalline platinum wire purchased from Aldrich (99.99% minimum purity certified reagent). The immersed end of the platinum electrode was curved at the end to a U-shape to mark for a constant surface area of the immersed part of the electrode. A silver/silver chloride was used as a quasi-reference electrode (QRE). The auxiliary electrode was a 0.5 mm polycrystalline platinum wire (Aldrich, certified 99.99% minimum purity). All reagents used were analytical grade and used as received from the suppliers without further purification. Sulfuric acid (95-97%) was supplied from Merck, L-ascorbic acid (99%) was purchased from AnalaR, potassium iodide was purchased from Sigma-Aldrich. Ultra-pure water, Millipore-MilliQ system was used for the preparation of all solutions. The N2 gas was a five grade, 99.999% minimum purity supplied from the International Jordanian Gases Company (Amman, Jordan).

Preparation of iodine-coated platinum electrode

The polycrystalline platinum electrode was cleaned with a freshly prepared chromic acid (H2CrO4), followed by rinsing with Millipore-Q water and sonicated for 10 minutes. After that, the platinum electrode was placed in contact with a supporting electrolyte solution of 0.5 M H2SO4 and conditioned between -0.25V and 1.3V until obtaining a reproducible cyclic voltammogram of a polycrystalline platinum electrode, which manifests the cleanliness of the electrode surface and electrochemical cell contents (Fig.1).

After cleaning the platinum electrode, the electrode was immersed in a supporting electrolyte containing 0.5 M H2SO4 + 0.01 M KI for five minutes under open-circuit conditions to complete the coating of the platinum electrode surface with iodine. Then the electrode was rinsed with water and 0.5M H2SO4 solution extensively. After the coating step, the electrode was cycled in a supporting electrolyte solution between -0.2V and +0.8V at a scan rate of 50 mV/s (Fig.1). The absence of oxygen and hydrogen adsorption/desorption features provides clear evidence for the complete coverage of the platinum electrode surface with a monolayer of iodine.

Sample preparation

The pharmaceutical formulation samples were purchased from local Jordanian pharmacies in the form of tablets and capsules. Three brands of pharmaceutical preparations were analyzed for their ascorbic acid content. The capsules of each sample were dissolved in 20 mL of 0.1 M KCl and sonicated for 10 min and left to equilibrate for 30 min. The solution was transferred to a 100 mL volumetric flask and filled to the mark with 0.1 M KCl. The solution was diluted to a concentration that matches the established calibration curve. A tablet of each sample was treated separately. The tablet of each brand was powdered using porcelain mortar and dissolved in 100.00 mL of the supporting electrolyte, 0.1 M KCl; the solution of the prepared samples was sonicated for 5 min and left to equilibrate for 5 min. A 5 mL aliquot of this solution was diluted to 50.00 mL with 0.1 M KCl to match the constructed calibration curve at different concentration ranges. A 10.00 mL of the diluted solution was placed in the electrochemical cell. The solution was bubbled with nitrogen gas (5G purity) and kept under a nitrogen gas atmosphere during the electrochemical experiment. The voltammetric analysis was conducted for ascorbic acid at the modified iodine-coated electrode within a potential window started at -0.2V and finished at 0.6V, where the adsorbed iodine is stable.

Results and discussions

Initially, a reproducible cyclic voltammogram for the polycrystalline platinum electrode, which indicates the cleanness of the electrochemical system, was obtained (Fig1-A). The process led to a successful coating process; the cyclic voltammogram of the iodine-coated platinum electrode between potential limits of -0.2V and 0.6V was displayed (Fig 1-B), where the adsorbed iodine was stable within this potential range. The complete absence of H2 and O2 oxidation-reduction features was the main indicator of a successful coated step.

Fig.1.

Cyclic voltammogramcurves of (A) polycrystalline platinum electrode and (B) the same electrode after adsorption of iodine from 0.01 M KI in 0.5 M H2SO4 solution

0121-4004-vitae-28-02-e7-gf1.png

The effect of varying supporting electrolytes on the anodic peak current of ascorbic acid oxidation was investigated. A 0.5 M H2SO4 (pH=0.3), phosphate buffer of pH=3.5, and 0.1 M KCl (pH=7) solution were used. As displayed in Figure 2, various oxidation peak current was obtained for ascorbic acid oxidation with different supporting electrolyte solutions. The highest oxidation peak current was obtained in 0.5 M KCl (the highest pH value), 45.47±0.09 mV. Therefore, 0.1 M KCl solution was considered a supporting electrolyte in the following study.

Fig.2.

Effectof type of supporting electrolyte on the electrochemical signal of ascorbic acid at iodine-coated platinum electrode, Ascorbic acid; 0.57mM. Supporting electrolytes: H2SO4 0.5M (pH=0.3); phosphate buffer, pH=3.5; KCl 0.1M (pH=7) n=3; scan rate, 50 mV/sec. Numbers above the bars = mean±SD.

0121-4004-vitae-28-02-e7-gf2.png

The effect of scan rate on the obtained anodic peak current of ascorbic acid was studied. As presented in Fig.3, there is a linear relationship between the square root of scan rate and oxidation peak current of ascorbic acid over the range of 10-100 mV/s, which suggested a diffusion-controlled irreversible oxidation process of ascorbic acid at the iodine-coated platinum electrode.

Fig 3.

a) Cyclic voltammograms of iodine-coated platinum electrode in 0.1 M KCl and 50ppm of ascorbic acid recorded at 10, 20, 50, and 100 mv/s. b) The least square line for the ascorbic acid oxidation peak current vs. the square root of scan rate.

0121-4004-vitae-28-02-e7-gf3.png

The obtained cyclic voltammograms for the iodine-coated platinum electrode in a series of ascorbic acid standard solutions show that the oxidation current increased steadily with ascorbic acid concentration (Fig.4). Three voltammograms were recorded for each standard solution. The anodic peak current was extracted for each cyclic voltammogram.

Fig 4

Cyclic voltammograms of iodine-coated platinum electrode in 0.1 M KCl solution containing 0.28, 0.57, 1.1, and 1.7 mM of ascorbic acid. All the scans were recorded at a scan rate of 50 mV/s.

0121-4004-vitae-28-02-e7-gf4.png

Plotting the anodic peak current variation against ascorbic acid concentration gave a straight and extended dynamic range with concentrations ranging between 2.84 µM - 5.68 mM The calibration curve displayed in Fig 5 shows that all the variability of the response data around its mean; R2=0.9969, and the calibration equation is given by

Where I represents the anodic peak current which attributed to the ascorbic acid oxidation as it is shown in the following equation:

The precision, that is, the repeatability of the method, was assessed by extracting the anodic peak current of the recorded cyclic voltammograms for a solution containing 0.28 mM ascorbic acid. The achieved coefficient of variation for 10 successive measurements was 1.36%, indicating the high precision of the developed method.

The limit of detection based on the formula LOD=3.3Ϭ/S, and the limit of quantitation based on the formula LOQ=10Ϭ/S, where Ϭ represents the blank signal (background current), and S means the sensitivity of the calibration curve was calculated. The estimated limits were 1.0 µM and 3.01 µM, respectively. Thus, acceptable sensitivity of the applied voltammetric method with high precision was obtained. Higher sensitivity can be achieved by applying a more sensitive technique like differential pulse voltammetry (DPV). However, differential pulse voltammetry was not attempted because cyclic voltammetry provides satisfactory sensitivity for ascorbic acid determination in pharmaceutical formulations.

Fig.5.

Anextended calibration curve shows the relationship between ascorbic acid concentration in ppm and the oxidation peak current measured from cyclic voltammograms for ascorbic acid in 0.1 M KCl at iodine-coated platinum electrode. Scan rate=50 mv*s-1.

0121-4004-vitae-28-02-e7-gf5.png

Potential interference

The influence of vitamins B1, B6, B1, folic acid, citric acid, sucrose, glucose, and zinc were investigated in order to verify the existence of matrix effects of vitamin C capsules and multivitamins tablets on ascorbic acid determination using cyclic voltammetry. The recorded cyclic voltammograms for each of these compounds show the absence of any electrochemical response of iodine-coated platinum electrodes toward these compounds. Figure 6 shows the recorded voltammograms for a solution of Multivitamin sample (control) and after each addition of a known concentration of the ascorbic acid standard solution. The result proved the absence of any possible interference with ascorbic acid despite the various components included in the Multivitamin sample, Vitamin E, B1, B2, B6, Folic acid, Pantothenic acid, Biotin, and Niacin.

Fig.6.

Cyclic voltammograms of iodine-coated platinum electrode in 0.1 M KCl containing Multivitamin tablet solution with addition of 0.1, 0.2, and 0.3 mM of ascorbic acid standard solution. Scan rate=50mV/s.

0121-4004-vitae-28-02-e7-gf6.jpg

Recovery

The recovery experiment can be taken as evidence for the absence of interference. The feasibility of the developed voltammetric method for ascorbic acid determination was tested for three pharmaceutical formulation samples. Ascorbic acid standards of known concentration, 50 ppm and 60 ppm, were spiked into samples of tablet solutions in order to evaluate the percentage recovery for each brand of ascorbic acid. As listed in Table 1, the recovery values were found between 98.93±2.78 and 99.98±5.20 for all samples of ascorbic acid brands, showing the appropriateness of the iodine-coated platinum electrode for the quantitative analysis of ascorbic acid in pharmaceutical formulations.

Table 1

Recoveries of ascorbic acid from spiked pharmaceutical formulations obtained by the developed method

% Recovery Detected ascorbic acid after addition (mM), Spiked ascorbic acid (mM) Content of ascorbic acid (mM) Samples
99.98±5.20 0.51 0.227 0.28 Vitamin C plus®
98.93±2.78 0.51 0.227 0.28 Multivit+®
99.80±2.50 0.787 0.341 0.568 Vitamin C 1000®

Pharmaceutical preparation analysis

The developed voltammetric method was applied to analyze ascorbic acid in three brands of the pharmaceutical formulation, multivitamin tablets, and two brands of vitamin C capsules (Vito + multivitamin, vitamin C plus, and vitamin C 1000). The standard addition method was applied to a diluted sample analysis to avoid the matrix effect. The evaluation of ascorbic acid concentration was found to be more suitable with the aid of a calibration graph. The results for the analysis of these pharmaceutical formulations with the developed voltammetric method are given in Table 2.

The obtained results by applying a cyclic voltammetry technique at iodine-coated platinum electrode were compared with the labeled values claimed by manufacturers. The data displayed in table 1 show that all nominal values are within the 95% confidence interval, which indicates the evident absence of errors in the results. The relative errors of the analysis of the three types of pharmaceutical formulations were lower than 5%, which attests to the accuracy of the developed method. The measured coefficient of variation values (0.55-2.19%) was considered obvious evidence of the precision of the developed method. The paired t-test was used to examine the significant difference at 95% confidence level between the labeled values and the obtained results determined by the developed voltammetric method. Comparing the calculated t value (0.0039) with the critical t value (4.30 at p=0.05) (33), it is shown that this result supported the null hypothesis and indicated no significant difference between the values determined by the voltammetric method and the nominal value obtained from manufacturers.

Table.2

Ascorbicacid (Vitamin C) content in pharmaceutical formulations collected from the Jordanian local pharmacies as determined by cyclic voltammetry at iodine-coated platinum electrode.

Pharmaceutical preparation Nominal mass(in mg of ascorbic acid/capsule or tablet) Average of determined mass(in mg) of ascorbic acid/capsule or tablet(n=3) Standard deviation 95% confidence limits Relative error Coefficient of variation
Vitamin C plus® 500 497.42 3.18 497.42± 7.90 0.516 0.64%
Multivitamin+® 60 60.32 1.32 60.31± 3.28 0.53 2.19%
Vitamin C1000® 1000 997.53 5.48 997.531± 13.61 0.247 0.55%

A comparison between the developed voltammetric method and some of the common analytical and voltammetric methods for ascorbic acid determination in terms of detection limit and the linear range was displayed in Table 2 3. As shown, the iodine-coated platinum electrode exhibited a lower detection limit than that of other voltammetric methods (29-31) (34-36). In contrast, the obtained linear range was convenient and extended compared to other voltammetric methods. Also, the developed method has the advantages of simplicity in sample preparations and analysis, side by side with a short time of analysis and the low price of instrumentations compared with other methods (Table 3).

Table. 3

A comparison of analytical performance of developed voltammetric method using iodine-coated platinum electrode with other analytical methods reported in literature

Method Linear range Detection Limit(ppm) Reference
HPLC 56.78 µM-0.57 mM 1.7x10-4 mM [5]
Spectrophotometry Method A:3.69 µM-63.59 µM Method B:2.9 µM-90.85 µM Method A=0.85 µM, Method B=1.1 µM [4]
Fluorometry 0.7 µM-6.02 µM 0.23 µM [13]
Cyclic voltammetry 0.01 mM-0.101 mM 1.76 µM [34]
Differential pulse voltammetry 19.0 µM-0.21 mM 19.0 µM [35]
Square-wave voltammetry - 1.87 µM [36]
Voltammetry 2.84 µM - 5.68 mM 0.96 µM This work

Conclusion

In this work, a successive use of an iodine-coated platinum electrode to determine ascorbic acid was achieved. The developed method excludes any sophisticated procedures. In contrast, it is considered an applicable method for simplicity of analysis procedures. The reported extended dynamic range 2.84x10-3 - 5.68 mM of ascorbic acid supports the applicability of the voltammetric method for ascorbic acid analysis in pharmaceutical products. Based on the recovery experiment, the absence of any interference from the other ingredients of pharmaceutical formulations is considered an evident indicator of the selectivity of the developed method. The statistical analysis of the results showed no significant difference between the values obtained from the voltammetric method and the labeled values claimed by the manufacturers.

Aknowledgment

The authors (Mohammad Amayreh & Wafa Hourani) is thankful to Dr.Mohammed Hourani from the Department of Chemistry, The University of Jordan, for his continuous support and encouragement and for providing the electrochemical workstation.

References

1. Pisoschi AM, Danet AF, Kalinowski S. Ascorbic acid determination in commercial fruit juice samples by cyclic voltammetry. J Autom Methods Manag Chem. 2008; 2008:1-8 https://doi.org/10.1155/2008/937651

AM Pisoschi AF Danet S Kalinowski Ascorbic acid determination in commercial fruit juice samples by cyclic voltammetryJ Autom Methods Manag Chem.2008200818https://doi.org/10.1155/2008/937651

2. Chambial S, Dwivedi S, Shukla KK, John PJ, Sharma P. Vitamin C in disease prevention and cure: An Overview. Indian Journal of Clinical Biochemistry.2013; 28(4), 314-328. https://doi.org/10.1007/s12291-013-0375-3

S Chambial S Dwivedi KK Shukla PJ John P Sharma Vitamin C in disease prevention and cure: An OverviewIndian Journal of Clinical Biochemistry2013284314328https://doi.org/10.1007/s12291-013-0375-3

3. Perry M, Page N, Manthey D, Zavitz J. Scurvy: Dietary Discretion in a developed country. Clinical Practice and Cases in Emergency Medicine. 2018; 2(2): 147-150. https://doi.org/10.5811/cpcem.2018.1.36860

M Perry N Page D Manthey J Zavitz Scurvy: Dietary Discretion in a developed country. Clinical Practice and Cases in Emergency Medicine201822147150https://doi.org/10.5811/cpcem.2018.1.36860

4. Naidu KA. Vitamin C in human health and disease is still a mystery? An overview. Nutrition Journal. 2003; 2:1-10. https://doi.org/10.1186/1475-2891-2-7.

KA. Naidu Vitamin C in human health and disease is still a mystery? An overviewNutrition Journal.20032110https://doi.org/10.1186/1475-2891-2-7

5. Hu L, Li L, Luo Z, Yang J, Liu W. Determination of trace vitamin C by ion-pair HPLC with UV detection in calcium gluconate and vitamin C compound oral solution. Journal of Chromatographic Science. 2012; 50(2):102-107. https://doi.org/10.1093/chromsci/bmr035

L Hu L Li Z Luo J Yang W Liu Determination of trace vitamin C by ion-pair HPLC with UV detection in calcium gluconate and vitamin C compound oral solutionJournal of Chromatographic Science2012502102107https://doi.org/10.1093/chromsci/bmr035

6. Gazdik Z, Zitka O, Petrlova J, Adam V, Zehnalek J, Horna A, Reznicek V, Beklova M, Kizek, R. Determination of vitamin C (Ascorbic Acid) using high performance liquid chromatography coupled with electrochemical detection. Sensors.2008; 8(11): 7097-7112. https://doi.org/10.3390/s8117097

Zitka O Gazdik Z Adam V Petrlova J Horna A Zehnalek J Beklova M Reznicek V R Kizek Determination of vitamin C (Ascorbic Acid) using high performance liquid chromatography coupled with electrochemical detectionSensors200881170977112https://doi.org/10.3390/s8117097

7. Shrestha N, Shrestha S, Bhattarai A. Determination of ascorbic acid in different citrus fruits of Kathmandu Valley. Journal of medical and Biological Science Research. 2016; 2(1):9-14.

N Shrestha S Shrestha A Bhattarai Determination of ascorbic acid in different citrus fruits of Kathmandu ValleyJournal of medical and Biological Science Research201621914

8. Anal PD, Shuchi, D. UV spectroscopic method for determination of vitamin C(ascorbic acid) content in different fruits in south Gujarat Region. International Journal of Environmental Sciences & Natural Resources. 2019; 21(2):41-44. https://doi.org/10.19080/IJESNR.2019.21.556056.

PD Anal D Shuchi UV spectroscopic method for determination of vitamin C(ascorbic acid) content in different fruits in south Gujarat RegionInternational Journal of Environmental Sciences & Natural Resources20192124144https://doi.org/10.19080/IJESNR.2019.21.556056

9. Mirsad S, Amra S. Spectrophotometric determination of L-ascorbic Acid in pharmaceutical based on its oxidation by potassium peroxymonosulfate and hydrogen Peroxide. Croatica Chemica Acta. 2015; 88:73-79. https://doi.org/10.5562/cca2551.

S Mirsad S Amra Spectrophotometric determination of L-ascorbic Acid in pharmaceutical based on its oxidation by potassium peroxymonosulfate and hydrogen PeroxideCroatica Chemica Acta2015887379https://doi.org/10.5562/cca2551

10. Zanini DJ, Silva MH, Aguiar-Oliveira E, Mazalli MR, Kamimura ES, Maldonado RR. Spectrophotometric analysis of vitamin C in different matrices utilizing potassium permanganate. European International Journal of Science and Technology.2018; 7(1):70-84.

DJ Zanini MH Silva E Aguiar-Oliveira MR Mazalli ES Kamimura RR Maldonado Spectrophotometric analysis of vitamin C in different matrices utilizing potassium permanganateEuropean International Journal of Science and Technology2018717084

11. Arya SP, Mahajan M, Jain P. Photometric methods for the determination of vitamin C. Analytical Sciences.1998; 14:889-895.

SP Arya M Mahajan P Jain Photometric methods for the determination of vitamin CAnalytical Sciences199814889895

12. Lau O-W, Luk S-F, Wong K-S. Determination of ascorbic acid in pharmaceuticals using direct ultraviolet spectrophotometry. Analyst. 1987; 112:1023-1025.

O-W Lau S-F Luk K-S Wong Determination of ascorbic acid in pharmaceuticals using direct ultraviolet spectrophotometryAnalyst198711210231025

13. Dilgin Y, Nisli G. Fluorimetric determination of ascorbic acid in vitamin C tablets using methylene blue. Chem. Pharm. Bull. 2015; 53:1251-1254. https://doi.org/10.1248/cpb.53.1251.

Y Dilgin G Nisli Fluorimetric determination of ascorbic acid in vitamin C tablets using methylene blueChem. Pharm. Bull.20155312511254https://doi.org/10.1248/cpb.53.1251

14. Ensafi AA, Rezaei B. Flow injection analysis determination of ascorbic acid with spectrofluorimetric detection. Analytical Letters. 1998; 31:333-342. https://doi.org/10.1080/00032719808002049.

AA Ensafi B Rezaei Flow injection analysis determination of ascorbic acid with spectrofluorimetric detectionAnalytical Letters199831333342https://doi.org/10.1080/00032719808002049

15. Gazdik Z, Zitka O, Petrlova J, Adam V, Zehnalek J, Horna A, Reznicek V, Beklova M, Kizek R. Determination of vitamin C(ascorbic acid) using high performance liquid chromatography coupled with electrochemical detection. Sensors .2008; 8(11):7097-7112. https://doi.org/ 10.3390/s8117097

Z Gazdik O Zitka J Petrlova V Adam J Zehnalek A Horna V Reznicek M Beklova R Kizek Determination of vitamin C(ascorbic acid) using high performance liquid chromatography coupled with electrochemical detectionSensors200881170977112https://doi.org/ 10.3390/s8117097

16. Clucu AA. Chemically modified electrodes in biosensing. J Biosens Bioelectron. 2014; 5(3):1-10. https://doi.org/10.4172/2155-6210.1000154

AA Clucu Chemically modified electrodes in biosensingJ Biosens Bioelectron201453110https://doi.org/10.4172/2155-6210.1000154

17. March G, Nguyen TD, Piro B. Modified electrodes used for electrochemical detection of metal ions in environmental analysis. Biosensor. 2015; 5(2):241-275. https://doi.org/10.3390/bios5020241.

G March TD Nguyen B Piro Modified electrodes used for electrochemical detection of metal ions in environmental analysisBiosensor201552241275https://doi.org/10.3390/bios5020241

18. Cox JA, Kulesza PJ. Oxidation and determination of nitrite at modified electrodes. Journal of Electroanalytical Chemistry and Interfacialelectrochemistry.1984; 175(1-2): 105-118. https://doi.org/10.1016/S0022-0728(84)80349-6

JA Cox PJ Kulesza Oxidation and determination of nitrite at modified electrodesJournal of Electroanalytical Chemistry and Interfacialelectrochemistry19841751-2105118https://doi.org/10.1016/S0022-0728(84)80349-6

19. Felter TE, Hubbard AT. L.E.E.D. and electrochemistry of iodine on Pt (100) and Pt (111) single-crystal surfaces. J. Electroanal. Chem. 1979, 100:473-491. https://doi.org/10.1016/S0022-0728(79)80179-5

TE Felter AT Hubbard L.E.E.D. and electrochemistry of iodine on Pt (100) and Pt (111) single-crystal surfacesJ. Electroanal. Chem.1979100473491https://doi.org/10.1016/S0022-0728(79)80179-5

20. Shu ZX, Bruckenstein S. Iodine Adsorption Studies at Platinum. J. Electroanal. Chem. 1991, 317:263-277. https://doi.org/10.1016/0022-0728(91)85019-L

ZX Shu S Bruckenstein Iodine Adsorption Studies at PlatinumJ. Electroanal. Chem.1991317263277https://doi.org/10.1016/0022-0728(91)85019-L

21. Mebrahtu T, Rodriguez JF, Bravo BG, Soriaga MP. Hydrogenative/cathodic stripping of iodine chemisorbed on smooth polycrystalline platinum electrode. J. Electroanal. Chem. 1987, 219, 327-333. https://doi.org/10.1016/0022-0728(87)85050-7

T Mebrahtu JF Rodriguez BG Bravo MP Soriaga Hydrogenative/cathodic stripping of iodine chemisorbed on smooth polycrystalline platinum electrodeJ. Electroanal. Chem.1987219327333https://doi.org/10.1016/0022-0728(87)85050-7

22. Thomas AE, Wieckowski A. Surface diffusion limited desorption of iodine on a platinum electrode?. Journal of Electroanalytical Chemistry.1995; 399:207-212. https://doi.org/10.1016/0022-0728(95)04226-1

AE Thomas A Wieckowski Surface diffusion limited desorption of iodine on a platinum electrode?Journal of Electroanalytical Chemistry1995399207212https://doi.org/10.1016/0022-0728(95)04226-1

23. Podlovchenko BI, Kolyadko EA. Adsorption of carbon monoxide on platinized platinum electrode with preadsorbed iodine and iodide Anions. Russian Journal of Electrochemistry. 2003; 39: 823-827.

BI Podlovchenko EA Kolyadko Adsorption of carbon monoxide on platinized platinum electrode with preadsorbed iodine and iodide AnionsRussian Journal of Electrochemistry200339823827

24. Hourani MK. Determination of silver (I) by cyclic voltammetry at iodine-coated electrodes. Analyst. 1994; 119: 1975-1978. https://doi.org/10.1039/AN9941901975

MK Hourani Determination of silver (I) by cyclic voltammetry at iodine-coated electrodesAnalyst199411919751978https://doi.org/10.1039/AN9941901975

25. Lane RF, Hubbard AT, Fukunaga K, Blanchard RJ. Brain catecholamines: detection in vivo by means of differential pulse voltammetry at surface-modified platinum electrodes. Brain Research. 1976; 114(2): 346-352. https://doi.org/10.1016/0006-8993(76)90678-8

RF Lane AT Hubbard K Fukunaga RJ Blanchard Brain catecholamines: detection in vivo by means of differential pulse voltammetry at surface-modified platinum electrodesBrain Research19761142346352https://doi.org/10.1016/0006-8993(76)90678-8

26. Hourani M, Jarar A, Arar S. Atmospheric SO2 determination by voltammetric analysis at an iodine-coated platinum electrode. Electroanalysis.1999; 11(9):637-640. https://doi.org/10.1002/(SICI)1521-4109(199907)11:9<637::AID-ELAN637>3.0.CO;2-R

M Hourani A Jarar S Arar Atmospheric SO2 determination by voltammetric analysis at an iodine-coated platinum electrodeElectroanalysis1999119637640https://doi.org/10.1002/(SICI)1521-4109(199907)11:9<637::AID-ELAN637>3.0.CO;2-R

27. Hourani Mk, Hijaz B. Voltammetric Analysis of Hydroquinone and Catechol at Iodine-Coated Polycrystalline Platinum Electrode. Journal of Natural and engineering Science.2014; 8(2): 25-29.

Hourani Mk B Hijaz Voltammetric Analysis of Hydroquinone and Catechol at Iodine-Coated Polycrystalline Platinum ElectrodeJournal of Natural and engineering Science2014822529

28. Amayreh M, Hourani MK. Determination of iron in dietary supplements by voltammetric analysis at an iodine-coated polycrystalline platinum electrode. Int.J. Electrochem. Sci. 2018; 13: 975-983. https://doi.org/10.20964/2018.01.81

M Amayreh MK Hourani Determination of iron in dietary supplements by voltammetric analysis at an iodine-coated polycrystalline platinum electrodeInt.J. Electrochem. Sci.201813975983https://doi.org/10.20964/2018.01.81

29. Amayreh M, Hournai MK. Direct determination of hemoglobin in blood using iodine-coated platinum polycrystalline electrode. Analytical and Bioanalytical Chemistry Research. 2019; 6(1): 59-68. https://doi.org/10.22036/ABCR.2018.125953.1198.

M Amayreh MK Hournai Direct determination of hemoglobin in blood using iodine-coated platinum polycrystalline electrodeAnalytical and Bioanalytical Chemistry Research2019615968https://doi.org/10.22036/ABCR.2018.125953.1198

30. Hournai MK, Amayreh M, Hourani W. A Voltammetric sensor based on iodine-coated platinum electrode for determination of iron in blood serum. Anal.Bioanal.Electrochem. 2018; 10(12):1620-1628.

MK Hournai M Amayreh W Hourani A Voltammetric sensor based on iodine-coated platinum electrode for determination of iron in blood serumAnal.Bioanal.Electrochem.2018101216201628

31. Amayreh M, Hourani M. Determination of Iron in Spinach Using Sweep Voltammetry at Iodine-Coated Platinum rotating Disk Electrode. Journal of AOAC International. 2019; 102(2):666-668. https://doi.org/10.5740/jaoacint.18-0267.

M Amayreh M Hourani Determination of Iron in Spinach Using Sweep Voltammetry at Iodine-Coated Platinum rotating Disk ElectrodeJournal of AOAC International20191022666668https://doi.org/10.5740/jaoacint.18-0267

32. Amayreh M, Hourani W, Hourani MK. Anodic Stripping Voltammetric Determination of Copper in Multivitamin-Mineral Formulations using Iodine-Coated Platinum Electrode. Methods Objects Chem. Anal. 2021,16(1),48-56. https://doi.org/10.17721/moca.2021.48-56.

M Amayreh W Hourani MK Hourani Anodic Stripping Voltammetric Determination of Copper in Multivitamin-Mineral Formulations using Iodine-Coated Platinum ElectrodeMethods Objects Chem. Anal.20211614856https://doi.org/10.17721/moca.2021.48-56

33. Miller JN, Miller JC Statistics and Chemometrics for Analytical Chemistry.6th ed. Pearson, England;2010.43-45p.

JN Miller JC Miller Statistics and Chemometrics for Analytical Chemistry6PearsonEngland20104345

34. Bitew Z, Amare M. Electrochemical determination of ascorbic acid in pharmaceutical tablets using carbon paste electrode. Organic & Medicinal Chemistry International Journal. 2019; 8(5):1-9. https://doi.org/10.19080/OMCIJ.2019.08.555749

Z Bitew M Amare Electrochemical determination of ascorbic acid in pharmaceutical tablets using carbon paste electrodeOrganic & Medicinal Chemistry International Journal20198519https://doi.org/10.19080/OMCIJ.2019.08.555749

35. Lourenção BC, Medeiros RA, Rocha-Filho RC, Fatibello-Filho O, Simultaneous differential pulse voltammetric determination of ascorbic acid and caffeine in pharmaceutical formulations using a boron-doped diamond electrode. Electroanalysis .2010; 22(15):1717-1723. https://doi.org/10.1002/elan.200900612

BC Lourenção RA Medeiros RC Rocha-Filho O Fatibello-Filho Simultaneous differential pulse voltammetric determination of ascorbic acid and caffeine in pharmaceutical formulations using a boron-doped diamond electrodeElectroanalysis2010221517171723https://doi.org/10.1002/elan.200900612

36. Vedenyapina MD, Kazakova MM, Skundin AM. Voltammetric determination of ascorbic acid in pharmaceutical formulation on a boron doped diamond electrode. Russ. J. Phys. Chem. 2019; 93: 1178-1181. https://doi.org/10.1134/S0036024419060335

MD Vedenyapina MM Kazakova AM Skundin Voltammetric determination of ascorbic acid in pharmaceutical formulation on a boron doped diamond electrodeRuss. J. Phys. Chem.20199311781181https://doi.org/10.1134/S0036024419060335